• <style id="kcfgx"><mark id="kcfgx"></mark></style>

      Welcome to the website of Baoji Zhongheng Instrument Co., Ltd. Tel:+86-917-3512881
      Differential pressure transmitter Single crystal silicon pressure transmitter Electromagnetic flowmeter Capacitive pressure transmitter High precision pressure transmitter Intelligent pressure switch Diffused silicon pressure transmitter Flow switch Intelligent pressure transfer controller Thermometer

      Pressure transmitter


      BXY6001 Differential Pressure

      Product code

       

      Overview

      Product quotationProduct consulting

      Product introduction

      BXY6001- Differential Pressure Transmitter
      The BXY6001 differential pressure transmitter is used to measure the liquid level, density, pressure, and flow rate of liquids, gases, or vapors, and then convert it into a 4-20mADC HART current signal output. BXY6001 can also communicate with RST375 handheld terminal or RSM100 Modem for parameter setting, process monitoring, etc.
      Standard specifications
      (Calibrate the range based on the standard zero point, stainless steel 316L diaphragm, filled with silicone oil)
      1 Performance specifications
      Reference accuracy for range adjustment
      (Including linearity, backlash, and repeatability from zero)
      ± 0.075%
      If TD>10 (TD=maximum range/adjustable range), it is:
      ± (0.0075 × TD)%
      The square root output accuracy is 1.5 times the linear reference accuracy mentioned above
      Environmental temperature impact


      Range code

      -Total impact from 20 ℃ to 65 ℃

      A

      ±(0.45×TD+0.25)%×Span

      B

      ±(0.30×TD+0.20)%×Span

      C/D/E

      ±(0.20×TD+0.10)%×Span

      Range code

      -Total impact of 40 ℃~-20 ℃ and 65 ℃~85 ℃

      A

      ±(0.45×TD+0.25)%×Span

      B

      ±(0.30×TD+0.20)%×Span

      C/D/E

      ±(0.20×TD+0.10)%×Span

      Over range impact: ± 0.075% × Span
      Static pressure effect


      Range code

      influence quantity

      A

      ±(0.15%URL+0.10%Span)/4MPa

      B

      ±(0.10%URL+0.075%Span)/16MPa

      C/D/E

      ±(0.05%URL+0.05%Span)/16MPa

      Overvoltage effect

      Range code

      influence quantity

      A

      ±0.2%×Span/4MPa

      B

      ±0.2%×Span/16MPa

      C/D/E

      ±0.1%×Span/16MPa

      Long term stability

      Range code

      influence quantity

      A

      ±0.5%×Span/1year

      B

      ±0.2%×Span/1year

      C/D/E

      ±0.1%×Span/1year

      Power impact: ± 0.001%/10V (12-42V DC), negligible.
      2 Functional specifications
      Range and range



      Range/
      range


      kPa

      mbar

      A

      Range

      0.1~1

      1~10

      Range

      -1~1

      -10~10

      B

      Range

      0.2~6

      2~60

      Range

      -6~6

      -60~60

      C

      Range

      0.4~40

      4~400

      Range

      -40~40

      -400~400

      D

      Range

      2.5~250

      25~2500

      Range

      -250~250

      -2500~2500

      E

      Range

      20~2000

      0.2~20 bar

      Range

      -500~2000

      -5~20bar

      Range limit
      Within the upper and lower limits of the range, it can be adjusted arbitrarily.
      It is recommended to select range codes with the lowest possible range ratio to optimize performance characteristics.
      Zero point setting
      The zero point and range can be adjusted to any value within the measurement range in the table, as long as the calibrated range is ≥ the minimum range
      Installation position impact
      The change in installation position parallel to the film surface will not cause zero drift. If the installation position changes by more than 90 ° to the film surface, there will be zero drift within the range of<0.4kPa, which can be corrected by adjusting zero. No range impact.
      output
      2-wire system, 4-20mADC, selectable for digital communication with HART output, with options for linear or square root output.
      Output signal limit: Imin=3.9mA, Imax=20.mA
      Alarm current
      Low alarm mode (minimum): 3.7 mA
      High alarm mode (maximum): 21 mA
      No alarm mode (hold): maintain the effective current value before the fault
      Alarm current standard setting: high alarm mode
      response time
      The damping constant of the amplifier component is 0.1s; The sensor time constant is between 0.1 and 1.6 seconds, depending on the range and range ratio. The additional adjustable time constant is 0.1~60s.
      The impact on nonlinear output (such as square root function) depends on that function and can be calculated based on it.
      Preheating time:<15s
      ambient temperature
      -40-85 ℃
      When equipped with LCD display and fluororubber sealing ring, -20~65 ℃
      Storage temperature/transportation temperature
      -50-85 ℃; With LCD display: -40~85 ℃
      working pressure
      The rated working pressure is divided into three levels: 16MPa, 25MPa, and 40MPa
      Static pressure limit
      From an absolute pressure of 3.5kPa to the rated pressure, the protective pressure can be greater than 1.5 times the rated pressure and applied to both sides of the transmitter.
      Unidirectional overload limit: Unidirectional overload can reach rated pressure
      Electromagnetic compatibility (EMC)
      Refer to the "Electromagnetic Compatibility Schedule" on the next page
      3 Installation
      Power supply and load conditions
      The power supply voltage is 24V, R ≤ (Us-12V)/Imax k Ω
      Where Imax=23 mA
      Maximum power supply voltage: 42VDC
      Minimum power supply voltage: 12VDC, 15VDC (backlight LCD display)
      Digital communication load range: 250~600 Ω
      Electrical connections
      M20X1.5 cable sealing buckle and wiring terminal are suitable for wires ranging from 0.5 to 2.5mm2.
      Process Connection
      The two ends of the process connection flange have NPT 1/4 and UNF 7/16 internal threads.
      4 Physical specifications
      material quality
      Measurement capsule: stainless steel 316L
      Diaphragm: stainless steel 316L, Hastelloy C
      Process flange: stainless steel 304
      Nut and bolt: stainless steel (A4)
      Filling fluid: silicone oil
      Sealing ring: nitrile rubber (NBR), fluororubber

      Note: (1) A performance level description: During testing, the performance is normal within the technical specification limits.
      (2) B performance level description: During testing, the function or performance is temporarily reduced or lost, but it can recover on its own, and the actual operating status, storage, and data remain unchanged.
      External dimension unit (mm)
      Horizontal piping connection method (side) Horizontal piping connection method (front)

      墻裝連接方式 垂直配管連接方式

      5電氣連接圖